Detection:
Conditions: Column conditioning: 3 mL methanol, then 3 x 2 mL of buffer solution (see above); Do not let run the column dry!
Sample aspiration: 2 x 5 mL of the prepared sample is passed through the column by vacuum (flow rate of 1–2 mL/min)
Washing: With 3 mL buffer solution (see above) are washed impurities
Elution: Elute the sweeteners with 2 mL methanol (flow rate: 1 mL/min); wait 10 min before applying a second portion of 2 mL of methanol in the same 5 mL test tube.
Evaporate the solvent to 3 mL under a stream of nitrogen at ambient temperature.